Prodigy Autocode Manuel du propriétaire Page 32

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 76
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 31
PRODIGY
AutoCode® Blood Glucose Meter PRODIGY
AutoCode® Blood Glucose Meter
Owner’s Manual Owner’s Manual
Probable Cause Actions
Error in performing Read the instructions
the test. thoroughly and repeat
the test again.
Control solution bottle Shake the control
not shaken well. solution bottle vigorously
and repeat the test again.
Expired or Check the expiration
contaminated date and the discard date
control solution. of the control solution.
Control solution that Control solution
is too warm or too cold.
should come
to room temperature
(less than 86°F/30°C)
before testing.
Test strip deterioration. Please repeat the test
with a new test strip.
v The control solution test is out of
range.
62
Probable Cause Actions
Meter malfunction. Please contact
Customer Care at
1.800.243.2636 for
assistance.
v The control solution test is out of
range.
63
Vue de la page 31
1 2 ... 27 28 29 30 31 32 33 34 35 36 37 ... 75 76

Commentaires sur ces manuels

Pas de commentaire